Mobile application-based oximetry: a potential toolfor appropriate referral of patients with respiratory symptoms examined via telemedicine

HumanInsight Mobile application-based oximetry: a potential toolfor appropriate referral of patients with respiratory symptoms examined via telemedicine

Einstein (Sao Paulo). 2023 Jan 6;21:eAO0025. doi: 10.31744/einstein_journal/2023AO0025. eCollection 2023.

ABSTRACT

OBJECTIVE: To calculate the positive likelihood ratio to determine whether telemedicine is able to optimize referral to the emergency department.

METHODS: Unicenter study with 182 consecutive patients admitted to Hospital Israelita Albert Einstein due to respiratory symptoms. All patients were submitted to oxygen saturation measurement using the standard method Welch Allyn finger device vital sign monitor and a 2-minute evaluation (Binah.ai mobile application). The reproducibility of oxygen saturation measurements made with both methods was investigated using interclass correlation coefficients and analysis of dispersion. Bland-Altman plots were constructed and kappa concordance coefficients used to examine data normality. Accuracy was also estimated.

RESULTS: Oxygen saturation measurement differences between methods were ≤2% in more than 85% of cases. The mean difference (bias) between methods was near zero (0.835; Bland-Altman analysis). Oxygen saturation measurements made using the Binah.ai mobile application had an average ability to detect patients with altered oxygen saturation levels compared to the conventional method (ROC analysis). The positive likelihood ratio of the mobile application was 6.23.

CONCLUSION: Mobile applications for oxygen saturation measurement are accessible user-friendly tools with moderate impact on clinical telemedicine evaluation of patients with respiratory symptoms, and may optimize referral to the emergency department.

PMID:36629679 | DOI:10.31744/einstein_journal/2023AO0025

Powered by WPeMatico

P.IVA 08738511214
Privacy Policy
Cookie Policy

Sede Legale
Viale Campi Flegrei 55
80124 - Napoli

Sede Operativa
Via G.Porzio 4
Centro Direzionale G1
80143 - Napoli

ISO9001
AI 4394
© Copyright 2022 - Humaninsight Srls - All Rights Reserved
Privacy Policy | Cookie Policy
envelopephone-handsetmap-marker linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram