Comparison of Novel Telemonitoring System Using the Single-lead Electrocardiogram Patch With Conventional Telemetry System

HumanInsight Comparison of Novel Telemonitoring System Using the Single-lead Electrocardiogram Patch With Conventional Telemetry System

Korean Circ J. 2024 Mar;54(3):140-153. doi: 10.4070/kcj.2023.0252.

ABSTRACT

BACKGROUND AND OBJECTIVES: Although a single-lead electrocardiogram (ECG) patch may provide advantages for detecting arrhythmias in outpatient settings owing to user convenience, its comparative effectiveness for real-time telemonitoring in inpatient settings remains unclear. We aimed to compare a novel telemonitoring system using a single-lead ECG patch with a conventional telemonitoring system in an inpatient setting.

METHODS: This was a single-center, prospective cohort study. Patients admitted to the cardiology unit for arrhythmia treatment who required a wireless ECG telemonitoring system were enrolled. A single-lead ECG patch and conventional telemetry were applied simultaneously in hospitalized patients for over 24 hours for real-time telemonitoring. The basic ECG parameters, arrhythmia episodes, and signal loss or noise were compared between the 2 systems.

RESULTS: Eighty participants (mean age 62±10 years, 76.3% male) were enrolled. The three most common indications for ECG telemonitoring were atrial fibrillation (66.3%), sick sinus syndrome (12.5%), and atrioventricular block (10.0%). The intra-class correlation coefficients for detecting the number of total beats, atrial and ventricular premature complexes, maximal, average, and minimal heart rates, and pauses were all over 0.9 with p values for reliability <0.001. Compared to a conventional system, a novel system demonstrated significantly lower signal noise (median 0.3% [0.1-1.6%] vs. 2.4% [1.4-3.7%], p<0.001) and fewer episodes of signal loss (median 22 [2-53] vs. 64 [22-112] episodes, p=0.002).

CONCLUSIONS: The novel telemonitoring system using a single-lead ECG patch offers performance comparable to that of a conventional system while significantly reducing signal loss and noise.

TRIAL REGISTRATION: Clinical Research Information Service Identifier: KCT0008176.

PMID:38506104 | DOI:10.4070/kcj.2023.0252

Powered by WPeMatico

P.IVA 08738511214
Privacy Policy
Cookie Policy

Sede Legale
Viale Campi Flegrei 55
80124 - Napoli

Sede Operativa
Via G.Porzio 4
Centro Direzionale G1
80143 - Napoli

ISO9001
AI 4394
© Copyright 2022 - Humaninsight Srls - All Rights Reserved
Privacy Policy | Cookie Policy
envelopephone-handsetmap-marker linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram