Accuracy and Reliability of Smartphone Self-Test Audiometry in Community Clinics in Low Income Settings: A Comparative Study.

Accuracy and Reliability of Smartphone Self-Test Audiometry in Community Clinics in Low Income Settings: A Comparative Study.

Accuracy and Reliability of Smartphone Self-Test Audiometry in Community Clinics in Low Income Settings: A Comparative Study.

Ann Otol Rhinol Laryngol. 2020 Jan 22;:3489420902162

Authors: Sandström J, Swanepoel D, Laurent C, Umefjord G, Lundberg T

Abstract
BACKGROUND: There is a lack of hearing health care globally, and tele-audiology and mobile technologies have been proposed as important strategies to reduce the shortfall.
OBJECTIVES: To investigate the accuracy and reliability of smartphone self-test audiometry in adults, in community clinics in low-income settings.
METHODS: A prospective, intra-individual, repeated measurements design was used. Sixty-three adult participants (mean age 52 years, range 20-88 years) were recruited from ENT and primary health care clinics in a low-income community in Tshwane, South Africa. Air conduction hearing thresholds for octave frequencies 0.5 to 8 kHz collected with the smartphone self-test in non-sound treated environments were compared to those obtained by reference audiometry.
RESULTS: The overall mean difference between threshold seeking methods (ie, smartphone thresholds subtracted from reference) was -2.2 dB HL (n = 467 thresholds, P = 0.00). Agreement was within 10 dB HL for 80.1% (n = 467 thresholds) of all threshold comparisons. Sensitivity for detection hearing loss >40 dB HL in one ear was 90.6% (n = 84 ears), and specificity 94.2% (n = 84 ears).
CONCLUSION: Smartphone self-test audiometry can provide accurate and reliable air conduction hearing thresholds for adults in community clinics in low-income settings.

PMID: 31965808 [PubMed - as supplied by publisher]

Powered by WPeMatico

P.IVA 08738511214
Privacy Policy
Cookie Policy

Sede Legale
Viale Campi Flegrei 55
80124 - Napoli

Sede Operativa
Via G.Porzio 4
Centro Direzionale G1
80143 - Napoli

ISO9001
AI 4394
© Copyright 2022 - Humaninsight Srls - All Rights Reserved
Privacy Policy | Cookie Policy
envelopephone-handsetmap-marker linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram